NCT04299425 · OUTCOME

Number of 'False Positive' Tissues Excised by Surgeon

Evaluating Impact of NIRAF Detection for Identifying Parathyroid Glands During Parathyroidectomy · Source last updated 2024-05-13

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
NUMBER
Unit
false positives
Interval / dispersion
Not reported
Time frame
Immediate to 10 days after PTx procedure.

What was measured

Number of tissues that were excised by surgeon assumed to be parathyroid tissue, but is later validated as non-parathyroid tissue (false positive) by histology

Analysis population: Only the NIRAF Detection Technology + arm was included in the analysis to calculate the rate of false positives by the device. Since the NIRAF Detection Technology - did not use the device that arm was not included.

Groups in this outcome

NIRAF Detection Technology +

Parathyroid gland identification will be performed with PTeye using NIRAF detection technology as an adjunctive tool in patients who undergo parathyroidectomy (PTx). NIRAF Detection Technology: Near Infrared Autofluorescence (NIRAF) detection technology or 'PTeye' consists of a disposable fiber-optic probe that emits non-ionizing radiation from a NIR 785 nm laser source, and also transmits the resulting tissue NIRAF to a photo detector. The 785 nm laser source emits a maximum power of 20 mW. The device is FDA cleared for clinical use in general surgeries and dermatological use (Class 2 device). After surgeon identifies a potential parathyroid gland in the surgical field, the surgeon places the fiber-optic probe of PTeye on suspect tissue and presses the device foot-pedal to activate tissue NIRAF measurement. Auditory beep at high frequency with a Detection Ratio \> 1.2 is interpreted by device as the suspect tissue being positive for parathyroid.

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
NIRAF Detection Technology +80

Reported measurements

Source class 1
Values in false positives · Interval/dispersion: Not reported
GroupValueSpreadLowerUpperComment
NIRAF Detection Technology +19Not reportedNot reportedNot reportedNot reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        value
        19
denoms
  1. counts
    1. group Id
      OG000
      value
      80
    units
    Participants
description
Number of tissues that were excised by surgeon assumed to be parathyroid tissue, but is later validated as non-parathyroid tissue (false positive) by histology
groups
  1. description
    Parathyroid gland identification will be performed with PTeye using NIRAF detection technology as an adjunctive tool in patients who undergo parathyroidectomy (PTx). NIRAF Detection Technology: Near Infrared Autofluorescence (NIRAF) detection technology or 'PTeye' consists of a disposable fiber-optic probe that emits non-ionizing radiation from a NIR 785 nm laser source, and also transmits the resulting tissue NIRAF to a photo detector. The 785 nm laser source emits a maximum power of 20 mW. The device is FDA cleared for clinical use in general surgeries and dermatological use (Class 2 device). After surgeon identifies a potential parathyroid gland in the surgical field, the surgeon places the fiber-optic probe of PTeye on suspect tissue and presses the device foot-pedal to activate tissue NIRAF measurement. Auditory beep at high frequency with a Detection Ratio \> 1.2 is interpreted by device as the suspect tissue being positive for parathyroid.
    id
    OG000
    title
    NIRAF Detection Technology +
param Type
NUMBER
population Description
Only the NIRAF Detection Technology + arm was included in the analysis to calculate the rate of false positives by the device. Since the NIRAF Detection Technology - did not use the device that arm was not included.
reporting Status
POSTED
time Frame
Immediate to 10 days after PTx procedure.
title
Number of 'False Positive' Tissues Excised by Surgeon
type
SECONDARY
unit Of Measure
false positives