Skip to content
← Full study record

NCT04299425 · OUTCOME

Number of Frozen Sections Sent for Analysis

Evaluating Impact of NIRAF Detection for Identifying Parathyroid Glands During Parathyroidectomy · Source last updated 2024-05-13

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
NUMBER
Unit
frozen sections
Interval / dispersion
Not reported
Time frame
Immediate. During PTx procedure.

What was measured

Number of frozen sections sent for analysis during the procedure to confirm potential parathyroid tissue

Groups in this outcome

NIRAF Detection Technology +

Parathyroid gland identification will be performed with PTeye using NIRAF detection technology as an adjunctive tool in patients who undergo parathyroidectomy (PTx). NIRAF Detection Technology: Near Infrared Autofluorescence (NIRAF) detection technology or 'PTeye' consists of a disposable fiber-optic probe that emits non-ionizing radiation from a NIR 785 nm laser source, and also transmits the resulting tissue NIRAF to a photo detector. The 785 nm laser source emits a maximum power of 20 mW. The device is FDA cleared for clinical use in general surgeries and dermatological use (Class 2 device). After surgeon identifies a potential parathyroid gland in the surgical field, the surgeon places the fiber-optic probe of PTeye on suspect tissue and presses the device foot-pedal to activate tissue NIRAF measurement. Auditory beep at high frequency with a Detection Ratio \> 1.2 is interpreted by device as the suspect tissue being positive for parathyroid.

NIRAF Detection Technology -

Parathyroid gland identification will be performed with the naked eye of the surgeon without using PTeye - NIRAF detection technology in patients who undergo parathyroidectomy (PTx).

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
NIRAF Detection Technology +80
NIRAF Detection Technology -80

Reported measurements

Source class 1
Values in frozen sections · Interval/dispersion: Not reported
GroupValueSpreadLowerUpperComment
NIRAF Detection Technology +13Not reportedNot reportedNot reportedNot reported
NIRAF Detection Technology -45Not reportedNot reportedNot reportedNot reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        value
        13
      2. group Id
        OG001
        value
        45
denoms
  1. counts
    1. group Id
      OG000
      value
      80
    2. group Id
      OG001
      value
      80
    units
    Participants
description
Number of frozen sections sent for analysis during the procedure to confirm potential parathyroid tissue
groups
  1. description
    Parathyroid gland identification will be performed with PTeye using NIRAF detection technology as an adjunctive tool in patients who undergo parathyroidectomy (PTx). NIRAF Detection Technology: Near Infrared Autofluorescence (NIRAF) detection technology or 'PTeye' consists of a disposable fiber-optic probe that emits non-ionizing radiation from a NIR 785 nm laser source, and also transmits the resulting tissue NIRAF to a photo detector. The 785 nm laser source emits a maximum power of 20 mW. The device is FDA cleared for clinical use in general surgeries and dermatological use (Class 2 device). After surgeon identifies a potential parathyroid gland in the surgical field, the surgeon places the fiber-optic probe of PTeye on suspect tissue and presses the device foot-pedal to activate tissue NIRAF measurement. Auditory beep at high frequency with a Detection Ratio \> 1.2 is interpreted by device as the suspect tissue being positive for parathyroid.
    id
    OG000
    title
    NIRAF Detection Technology +
  2. description
    Parathyroid gland identification will be performed with the naked eye of the surgeon without using PTeye - NIRAF detection technology in patients who undergo parathyroidectomy (PTx).
    id
    OG001
    title
    NIRAF Detection Technology -
param Type
NUMBER
reporting Status
POSTED
time Frame
Immediate. During PTx procedure.
title
Number of Frozen Sections Sent for Analysis
type
SECONDARY
unit Of Measure
frozen sections

Download exact source JSON → · Snapshot ctgov-results-25a395029948b04d52b1253f

Preserved source evidence · Not medical advice