NCT04083378 · OUTCOME

Minimal Ablative Margin Assessment on Post-ablation Intraprocedural CT.

Software-Aided Imaging (Morfeus) for Confirming Tumor Coverage With Ablation in Patients With Liver Tumors, the COVER-ALL Study · Source last updated 2026-03-23

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
MEAN
Unit
mm
Interval / dispersion
Standard Deviation
Time frame
Visit 2 (baseline/ablation day)

What was measured

For control group, MAM was assessed by rigid co-registration with side-by-side juxtaposition to verify applicator placement and by visual landmark-based CT measurements on CT workstation (Siemens Healthineers; Forchheim, Germany). For experimental group, CT images were processed with a software-based method on a radiation treatment planning system (Raystation, RaySearch Laboratories; Stockholm, Sweden) with biomechanical deformable imaging registration coupled with AI-based autosegmentation. For experimental group, interventional radiologist (IR) was informed of ablation applicator position relative to tumor on non-contrast CT before thermal ablation energy delivery and of MAM results (including spatial location of suboptimal margins) generated by software-based assessment, with this information not disclosed to IR in control group. Software-based assessment was used to compare MAMs results between both study groups with a two-sample t-test.

Analysis population: The primary endpoint included all enrolled patients

Groups in this outcome

Experimental

Clinical impact of Morfeus software usage on patients undergoing percutaneous ablation of liver tumors

Control

Morfeus software was not used during percutaneous ablation of liver tumors

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
Experimental74
Control26

Reported measurements

Source class 1
Values in mm · Interval/dispersion: Standard Deviation
GroupValueSpreadLowerUpperComment
Experimental6.82.9Not reportedNot reportedNot reported
Control2.22.8Not reportedNot reportedNot reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        spread
        2.9
        value
        6.8
      2. group Id
        OG001
        spread
        2.8
        value
        2.2
denoms
  1. counts
    1. group Id
      OG000
      value
      74
    2. group Id
      OG001
      value
      26
    units
    Participants
description
For control group, MAM was assessed by rigid co-registration with side-by-side juxtaposition to verify applicator placement and by visual landmark-based CT measurements on CT workstation (Siemens Healthineers; Forchheim, Germany). For experimental group, CT images were processed with a software-based method on a radiation treatment planning system (Raystation, RaySearch Laboratories; Stockholm, Sweden) with biomechanical deformable imaging registration coupled with AI-based autosegmentation. For experimental group, interventional radiologist (IR) was informed of ablation applicator position relative to tumor on non-contrast CT before thermal ablation energy delivery and of MAM results (including spatial location of suboptimal margins) generated by software-based assessment, with this information not disclosed to IR in control group. Software-based assessment was used to compare MAMs results between both study groups with a two-sample t-test.
dispersion Type
Standard Deviation
groups
  1. description
    Clinical impact of Morfeus software usage on patients undergoing percutaneous ablation of liver tumors
    id
    OG000
    title
    Experimental
  2. description
    Morfeus software was not used during percutaneous ablation of liver tumors
    id
    OG001
    title
    Control
param Type
MEAN
population Description
The primary endpoint included all enrolled patients
reporting Status
POSTED
time Frame
Visit 2 (baseline/ablation day)
title
Minimal Ablative Margin Assessment on Post-ablation Intraprocedural CT.
type
PRIMARY
unit Of Measure