Skip to content
← Full study record

NCT03862131 · OUTCOME

Accuracy of MyoStrain® and CMR Standard Measurements (LV End Diastolic Volume Index, LV End Systolic Volume Index, LV Stroke Volume Index and Left Ventricular Ejection Fraction) to Detect Myocardial Dysfunction During Cancer Treatment

PROactive Evaluation of Function to Avoid CardioToxicity · Source last updated 2025-07-29

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
NUMBER
Unit
percent accuracy
Interval / dispersion
95% Confidence Interval
Time frame
Through 36 months

What was measured

Accuracy is defined as (true positives + true negatives) divided by the total number of cases * Myocardial dysfunction was defined as the occurrence of either clinical or subclinical cardiotoxicity. * Clinical cardiotoxicity was defined as an absolute change in LVEF \> 10% from baseline to below 53% combined with heart failure symptoms or elevation in NT pro BNP to above 199 pg/mL. * In patients with LVEF \< 53% at baseline, clinical cardiotoxicity was defined by a symptomatic drop in LVEF \> 10% or a HF hospitalization. * Subclinical CTX was defined as an asymptomatic patient with a greater than 10% decrease in LVEF, worsening GLS more than 15% from baseline, or new elevation in NT pro BNP \>199 pg/mL.

Groups in this outcome

MyoStrain® Unblinded Treatment Arm

* After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk Group * The unblinded treatment arm will enhance patient management by augmenting standard of care with serial MyoStrain® monitoring of the impact of cancer therapy on myocardial function. * Higher Risk unblinded patients will continue to undergo MyoStrain® MRI testing, regardless of study arm, at 1 month (+1 week), 3 months (+ 1 week), 6 months (+1 week), 12 months (+ 30 days), 24 months (+30 days), and 36 months (+30 days) after the baseline visit. * In addition to the MyoStrain® testing, patients will also be asked to complete a brief patient satisfaction questionnaire at each PROACT time point.

MyoStrain® Blinded Control Arm

* After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk group * The blinded control arm will provide investigators with LVEF and LVEDV/LVESV measurements, which are clinical, in conjunction with standard of care * Higher Risk blinded patients will continue to undergo MyoStrain® MRI testing, regardless of study arm, at 1 month (+1 week), 3 months (+ 1 week), 6 months (+1 week), 12 months (+ 30 days), 24 months (+30 days), and 36 months (+30 days) after the baseline visit. * In addition to the MyoStrain® testing, patients will also be asked to complete a brief patient satisfaction questionnaire at each PROACT time point.

Combined MyoStrain® Control Arm and Treatment Arm

* After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk Group. * After meeting eligibility criteria, patients will be further randomized to an unblinded arm (MyoStrain® measured and available to the treatment team) and a blinded arm (MyoStrain® measured but not available to the treatment team). This randomization will aid in an exploratory aim of testing the feasibility of MyoStrain® to guide cardioprotective therapy in the unblinded vs blinded arms. However, as all patients will have MyoStrain® measured, all patients (both treatment arms) will be combined for testing the primary outcome measure.

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
MyoStrain® Unblinded Treatment Arm25
MyoStrain® Blinded Control Arm18
Combined MyoStrain® Control Arm and Treatment Arm43

Reported measurements

MyoStrain®
Values in percent accuracy · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
MyoStrain® Unblinded Treatment Arm78.6Not reported71.885.4Not reported
MyoStrain® Blinded Control Arm85.1Not reported78.292.1Not reported
Combined MyoStrain® Control Arm and Treatment Arm75.1Not reported69.680.6Not reported
CMR LVEF
Values in percent accuracy · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
MyoStrain® Unblinded Treatment Arm87.1Not reported81.592.6Not reported
MyoStrain® Blinded Control Arm84.8Not reported77.891.9Not reported
Combined MyoStrain® Control Arm and Treatment Arm80.3Not reported75.285.3Not reported
CMR left ventricular end-diastolic indexed (LVEDVI)
Values in percent accuracy · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
MyoStrain® Unblinded Treatment Arm71.2Not reported63.778.7Not reported
MyoStrain® Blinded Control Arm54.5Not reported44.764.4Not reported
Combined MyoStrain® Control Arm and Treatment Arm59.7Not reported53.465.9Not reported
CMR left ventricular end-systolic volume index (LVESVI)
Values in percent accuracy · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
MyoStrain® Unblinded Treatment Arm89.2Not reported84.194.4Not reported
MyoStrain® Blinded Control Arm68.7Not reported59.677.8Not reported
Combined MyoStrain® Control Arm and Treatment Arm72.3Not reported66.678.0Not reported
CMR left ventricular stroke volume index (LVSVI)
Values in percent accuracy · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
MyoStrain® Unblinded Treatment Arm84.9Not reported78.990.8Not reported
MyoStrain® Blinded Control Arm50.5Not reported40.760.4Not reported
Combined MyoStrain® Control Arm and Treatment Arm82.8Not reported78.087.6Not reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        71.8
        upper Limit
        85.4
        value
        78.6
      2. group Id
        OG001
        lower Limit
        78.2
        upper Limit
        92.1
        value
        85.1
      3. group Id
        OG002
        lower Limit
        69.6
        upper Limit
        80.6
        value
        75.1
    title
    MyoStrain®
  2. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        81.5
        upper Limit
        92.6
        value
        87.1
      2. group Id
        OG001
        lower Limit
        77.8
        upper Limit
        91.9
        value
        84.8
      3. group Id
        OG002
        lower Limit
        75.2
        upper Limit
        85.3
        value
        80.3
    title
    CMR LVEF
  3. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        63.7
        upper Limit
        78.7
        value
        71.2
      2. group Id
        OG001
        lower Limit
        44.7
        upper Limit
        64.4
        value
        54.5
      3. group Id
        OG002
        lower Limit
        53.4
        upper Limit
        65.9
        value
        59.7
    title
    CMR left ventricular end-diastolic indexed (LVEDVI)
  4. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        84.1
        upper Limit
        94.4
        value
        89.2
      2. group Id
        OG001
        lower Limit
        59.6
        upper Limit
        77.8
        value
        68.7
      3. group Id
        OG002
        lower Limit
        66.6
        upper Limit
        78.0
        value
        72.3
    title
    CMR left ventricular end-systolic volume index (LVESVI)
  5. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        78.9
        upper Limit
        90.8
        value
        84.9
      2. group Id
        OG001
        lower Limit
        40.7
        upper Limit
        60.4
        value
        50.5
      3. group Id
        OG002
        lower Limit
        78.0
        upper Limit
        87.6
        value
        82.8
    title
    CMR left ventricular stroke volume index (LVSVI)
denoms
  1. counts
    1. group Id
      OG000
      value
      25
    2. group Id
      OG001
      value
      18
    3. group Id
      OG002
      value
      43
    units
    Participants
description
Accuracy is defined as (true positives + true negatives) divided by the total number of cases * Myocardial dysfunction was defined as the occurrence of either clinical or subclinical cardiotoxicity. * Clinical cardiotoxicity was defined as an absolute change in LVEF \> 10% from baseline to below 53% combined with heart failure symptoms or elevation in NT pro BNP to above 199 pg/mL. * In patients with LVEF \< 53% at baseline, clinical cardiotoxicity was defined by a symptomatic drop in LVEF \> 10% or a HF hospitalization. * Subclinical CTX was defined as an asymptomatic patient with a greater than 10% decrease in LVEF, worsening GLS more than 15% from baseline, or new elevation in NT pro BNP \>199 pg/mL.
dispersion Type
95% Confidence Interval
groups
  1. description
    * After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk Group * The unblinded treatment arm will enhance patient management by augmenting standard of care with serial MyoStrain® monitoring of the impact of cancer therapy on myocardial function. * Higher Risk unblinded patients will continue to undergo MyoStrain® MRI testing, regardless of study arm, at 1 month (+1 week), 3 months (+ 1 week), 6 months (+1 week), 12 months (+ 30 days), 24 months (+30 days), and 36 months (+30 days) after the baseline visit. * In addition to the MyoStrain® testing, patients will also be asked to complete a brief patient satisfaction questionnaire at each PROACT time point.
    id
    OG000
    title
    MyoStrain® Unblinded Treatment Arm
  2. description
    * After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk group * The blinded control arm will provide investigators with LVEF and LVEDV/LVESV measurements, which are clinical, in conjunction with standard of care * Higher Risk blinded patients will continue to undergo MyoStrain® MRI testing, regardless of study arm, at 1 month (+1 week), 3 months (+ 1 week), 6 months (+1 week), 12 months (+ 30 days), 24 months (+30 days), and 36 months (+30 days) after the baseline visit. * In addition to the MyoStrain® testing, patients will also be asked to complete a brief patient satisfaction questionnaire at each PROACT time point.
    id
    OG001
    title
    MyoStrain® Blinded Control Arm
  3. description
    * After consenting to the PROACT study, patients will undergo a baseline MRI to determine their risk stratification for the study. This baseline MyoStrain® MRI must demonstrate 2 or more segments measuring \>-10% or 9 or more segments \>-17% for entrance into the study as the Higher Risk Group. * After meeting eligibility criteria, patients will be further randomized to an unblinded arm (MyoStrain® measured and available to the treatment team) and a blinded arm (MyoStrain® measured but not available to the treatment team). This randomization will aid in an exploratory aim of testing the feasibility of MyoStrain® to guide cardioprotective therapy in the unblinded vs blinded arms. However, as all patients will have MyoStrain® measured, all patients (both treatment arms) will be combined for testing the primary outcome measure.
    id
    OG002
    title
    Combined MyoStrain® Control Arm and Treatment Arm
param Type
NUMBER
reporting Status
POSTED
time Frame
Through 36 months
title
Accuracy of MyoStrain® and CMR Standard Measurements (LV End Diastolic Volume Index, LV End Systolic Volume Index, LV Stroke Volume Index and Left Ventricular Ejection Fraction) to Detect Myocardial Dysfunction During Cancer Treatment
type
PRIMARY
unit Of Measure
percent accuracy

Download exact source JSON → · Snapshot ctgov-results-25a395029948b04d52b1253f

Preserved source evidence · Not medical advice