Skip to content
← Full study record

NCT03686215 · OUTCOME

Sensitivity of LUM Imaging System on Cancer in the Lumpectomy Cavity

Investigation of Novel Surgical Imaging for Tumor Excision · Source last updated 2023-08-24

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
NUMBER
Unit
percentage of tissues with cancer
Interval / dispersion
95% Confidence Interval
Time frame
Day 1, during surgery

What was measured

Sensitivity of LUM Imaging System on cancer was calculated as percentage of tissues with cancer that was removed from lumpectomy cavity due to positive detection by LUM Imaging System. The estimates of the instrument diagnostic sensitivity used Generalized Estimating Equations (GEE) approach. This method was used to analyze correlated data which was modeled with generalized linear model having binomial link function and including correlation cluster and matrix information (here within each subject). LUM Imaging System was used during the surgery, but the assessment of the outcomes was unknown until the pathology report was available.

Analysis population: modified intent to treat population (mITT) includes subjects injected with LUM015 and LUM Imaging was performed

Groups in this outcome

Device Intervention: LUM Imaging System Used During Surgery

The LUM Imaging Device was used to examine inside the lumpectomy cavity to determine if the dye indicated any areas that might contain residual tumor. If the imaging tested positive for the cancer cells remaining in the lumpectomy cavity, the surgeon would remove an additional piece of tissue. This process was continued until a negative reading from the device was obtained or a maximum of 2 shaves of additional tissue had been removed. Patients in this arm received the study drug, LUM015. Study Device Arm: LUM015 was administered 2 to 6 hours prior to surgery. The LUM Imaging device was used to assist in the removal of additional tumor tissue.

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
Device Intervention: LUM Imaging System Used During Surgery357

Analysis denominator · Tissue with cancer

GroupSource count
Device Intervention: LUM Imaging System Used During Surgery69

Reported measurements

Source class 1
Values in percentage of tissues with cancer · Interval/dispersion: 95% Confidence Interval
GroupValueSpreadLowerUpperComment
Device Intervention: LUM Imaging System Used During Surgery49.1Not reported36.461.9Not reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        36.4
        upper Limit
        61.9
        value
        49.1
denom Units Selected
Tissue with cancer
denoms
  1. counts
    1. group Id
      OG000
      value
      357
    units
    Participants
  2. counts
    1. group Id
      OG000
      value
      69
    units
    Tissue with cancer
description
Sensitivity of LUM Imaging System on cancer was calculated as percentage of tissues with cancer that was removed from lumpectomy cavity due to positive detection by LUM Imaging System. The estimates of the instrument diagnostic sensitivity used Generalized Estimating Equations (GEE) approach. This method was used to analyze correlated data which was modeled with generalized linear model having binomial link function and including correlation cluster and matrix information (here within each subject). LUM Imaging System was used during the surgery, but the assessment of the outcomes was unknown until the pathology report was available.
dispersion Type
95% Confidence Interval
groups
  1. description
    The LUM Imaging Device was used to examine inside the lumpectomy cavity to determine if the dye indicated any areas that might contain residual tumor. If the imaging tested positive for the cancer cells remaining in the lumpectomy cavity, the surgeon would remove an additional piece of tissue. This process was continued until a negative reading from the device was obtained or a maximum of 2 shaves of additional tissue had been removed. Patients in this arm received the study drug, LUM015. Study Device Arm: LUM015 was administered 2 to 6 hours prior to surgery. The LUM Imaging device was used to assist in the removal of additional tumor tissue.
    id
    OG000
    title
    Device Intervention: LUM Imaging System Used During Surgery
param Type
NUMBER
population Description
modified intent to treat population (mITT) includes subjects injected with LUM015 and LUM Imaging was performed
reporting Status
POSTED
time Frame
Day 1, during surgery
title
Sensitivity of LUM Imaging System on Cancer in the Lumpectomy Cavity
type
PRIMARY
type Units Analyzed
Tissue with cancer
unit Of Measure
percentage of tissues with cancer

Download exact source JSON → · Snapshot ctgov-results-191f4516d4760045304b24a0