Skip to content
← Full study record

NCT01292486 · OUTCOME

Platelet Collection Efficiency

Evaluation of the Spectra Optia® Mononuclear Cell Collection Procedure · Source last updated 2013-05-01

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Measure
MEDIAN
Unit
% of processed platelets collected
Interval / dispersion
Full Range
Time frame
up to 7 days

What was measured

Platelet contamination of the cell product was measured as the platelet collection efficiency, that is, as the percent of platelets processed that were collected.

Analysis population: Data to calculate platelet collection efficiency was available for 38 MNC collections on 24 of the 26 per protocol patients.

Groups in this outcome

All Per Protocol Patients

Multiple myeloma patients infused with autologous peripheral blood stem cells collected on the Spectra Optia Apheresis System.

Group identifiers and denominators belong to this outcome only. They may differ in another outcome or in safety reporting.

Analysis denominator · Participants

GroupSource count
All Per Protocol Patients24

Analysis denominator · MNC collections

GroupSource count
All Per Protocol Patients38

Reported measurements

Source class 1
Values in % of processed platelets collected · Interval/dispersion: Full Range
GroupValueSpreadLowerUpperComment
All Per Protocol Patients19Not reported1338Not reported
Complete source fields
classes
  1. categories
    1. measurements
      1. group Id
        OG000
        lower Limit
        13
        upper Limit
        38
        value
        19
denom Units Selected
Participants
denoms
  1. counts
    1. group Id
      OG000
      value
      24
    units
    Participants
  2. counts
    1. group Id
      OG000
      value
      38
    units
    MNC collections
description
Platelet contamination of the cell product was measured as the platelet collection efficiency, that is, as the percent of platelets processed that were collected.
dispersion Type
Full Range
groups
  1. description
    Multiple myeloma patients infused with autologous peripheral blood stem cells collected on the Spectra Optia Apheresis System.
    id
    OG000
    title
    All Per Protocol Patients
param Type
MEDIAN
population Description
Data to calculate platelet collection efficiency was available for 38 MNC collections on 24 of the 26 per protocol patients.
reporting Status
POSTED
time Frame
up to 7 days
title
Platelet Collection Efficiency
type
SECONDARY
type Units Analyzed
MNC collections
unit Of Measure
% of processed platelets collected

Download exact source JSON → · Snapshot ctgov-results-25a395029948b04d52b1253f

Preserved source evidence · Not medical advice