NCT01045057 · OUTCOME
Success Rate of Procedure
Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.
- Measure
- NUMBER
- Unit
- Nr part. with succesful insertions
- Interval / dispersion
- Not reported
- Time frame
- immediate observation during surgery
What was measured
As successful counted all successful procedures in which the tracheal flange of the voice prosthesis unfolded completely without help of additional tools. A success rate of 80% and higher was considered acceptable.
Analysis population: Number subjects based on optimal minimax two-stage design. Based on power calculation, 26 patients should be included. In first stage, 20/23 successful insertions are needed to continue. In second stage, with 23/26 successful insertions device is considered safe. Analysis is per protocol.
Groups in this outcome
Provox Vega Puncture Set
Patients with laryngeal cancer and a total laryngectomy in whom a TE puncture was made and a voice prosthesis was placed by means of the Provox Vega Puncture Set
| Group | Source count |
|---|---|
| Provox Vega Puncture Set | 27 |
Reported measurements
Source class 1
| Group | Value | Spread | Lower | Upper | Comment |
|---|---|---|---|---|---|
| Provox Vega Puncture Set | 23 | Not reported | Not reported | Not reported | Not reported |
Complete source fields
- classes
- categories
- measurements
- group Id
- OG000
- value
- 23
- denoms
- counts
- group Id
- OG000
- value
- 27
- units
- Participants
- description
- As successful counted all successful procedures in which the tracheal flange of the voice prosthesis unfolded completely without help of additional tools. A success rate of 80% and higher was considered acceptable.
- groups
- description
- Patients with laryngeal cancer and a total laryngectomy in whom a TE puncture was made and a voice prosthesis was placed by means of the Provox Vega Puncture Set
- id
- OG000
- title
- Provox Vega Puncture Set
- param Type
- NUMBER
- population Description
- Number subjects based on optimal minimax two-stage design. Based on power calculation, 26 patients should be included. In first stage, 20/23 successful insertions are needed to continue. In second stage, with 23/26 successful insertions device is considered safe. Analysis is per protocol.
- reporting Status
- POSTED
- time Frame
- immediate observation during surgery
- title
- Success Rate of Procedure
- type
- PRIMARY
- unit Of Measure
- Nr part. with succesful insertions
Download exact source JSON → · Snapshot ctgov-results-25a395029948b04d52b1253f