Clinical Feasibility of New Tracheoesophageal Puncture Set · Source last updated 2021-06-11

Results reported by the registry submitting party. Population, time horizon, units and missing values must be read together.

Adverse events occurred during the study; they are not necessarily caused by treatment. Serious events are not interchangeable with severe/grade 3–4 events. These safety populations may differ from efficacy populations.

Collection time frame
During one month after the surgery
Other-event reporting threshold
0%

all adverse events, including expected/non-related events were recorded

Participants affected / participants at risk, as reported
Safety groupSerious eventsOther eventsDeaths
Provox Vega Puncture Set

Group of larynx cancer patients undergoing laryngectomy whereby the puncture and the placement of the voice prosthesis is done with the Provox Vega Puncture set

0 / 278 / 27Not reported / Not reported

Group identifiers: EG000: Provox Vega Puncture Set

Serious adverse events

Event terms and counts are source-reported; expand to read more. A reporting threshold may omit less frequent events.

    Other reported adverse events

    Event terms and counts are source-reported; expand to read more. A reporting threshold may omit less frequent events.

    1. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      In all 3 patients known significant risk factors were present that are deemed to have caused the fistula. As such, these complications are not considered related to device/procedure. Fistula occurrence rate was within previously reported ranges.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        3
        num At Risk
        27
        num Events
        3
      term
      Pharyngocutaneous Fistula
    2. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Posterior pharyngeal wall was damaged during dilatation before the puncture took place. Non-device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Damage posterior pharyngeal wall
    3. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Puncture penetrated posterior esophageal wall because window of device used for pharynx protection (not part of subject device)could not be located by palpation or visualization. Event did not cause any harm to patient. Event is not device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Damage posterior esophageal wall
    4. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Voice prosthesis retracted while cleaning it, and was repositioned. No patient harm. Event is related to patient anatomy, not related to device/procedure.Occurred in home situation, can occur when voice prosthesis (e.g. due to edema) fits too tight.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Retracted voice prosthesis
    5. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Patient developed tracheitis, caused by COPD and the related use of Prednisone and inhalation medications. Patient received antibiotics and was fine. Event was unrelated to device.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Tracheitis
    6. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Wound healing problems considered to be caused by pre-existing risk factors (Radiotherapy,prednisone use)and surgical risk factor of bilateral neck dissection.Not device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Wound dehiscence
    Complete source fields
    description
    all adverse events, including expected/non-related events were recorded
    event Groups
    1. description
      Group of larynx cancer patients undergoing laryngectomy whereby the puncture and the placement of the voice prosthesis is done with the Provox Vega Puncture set
      id
      EG000
      other Num Affected
      8
      other Num At Risk
      27
      serious Num Affected
      0
      serious Num At Risk
      27
      title
      Provox Vega Puncture Set
    frequency Threshold
    0
    other Events
    1. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      In all 3 patients known significant risk factors were present that are deemed to have caused the fistula. As such, these complications are not considered related to device/procedure. Fistula occurrence rate was within previously reported ranges.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        3
        num At Risk
        27
        num Events
        3
      term
      Pharyngocutaneous Fistula
    2. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Posterior pharyngeal wall was damaged during dilatation before the puncture took place. Non-device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Damage posterior pharyngeal wall
    3. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Puncture penetrated posterior esophageal wall because window of device used for pharynx protection (not part of subject device)could not be located by palpation or visualization. Event did not cause any harm to patient. Event is not device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Damage posterior esophageal wall
    4. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Voice prosthesis retracted while cleaning it, and was repositioned. No patient harm. Event is related to patient anatomy, not related to device/procedure.Occurred in home situation, can occur when voice prosthesis (e.g. due to edema) fits too tight.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Retracted voice prosthesis
    5. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Patient developed tracheitis, caused by COPD and the related use of Prednisone and inhalation medications. Patient received antibiotics and was fine. Event was unrelated to device.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Tracheitis
    6. assessment Type
      NON_SYSTEMATIC_ASSESSMENT
      notes
      Wound healing problems considered to be caused by pre-existing risk factors (Radiotherapy,prednisone use)and surgical risk factor of bilateral neck dissection.Not device related.
      organ System
      Surgical and medical procedures
      stats
      1. group Id
        EG000
        num Affected
        1
        num At Risk
        27
        num Events
        1
      term
      Wound dehiscence
    time Frame
    During one month after the surgery

    Download exact source JSON → · Snapshot ctgov-results-25a395029948b04d52b1253f